Skip to main content

Table 1 Comparison of software-generated and visual positive/negative classification

From: Performance analysis of automated evaluation of antinuclear antibody indirect immunofluorescent tests in a routine setting

 

Visual evaluation (observer 1)

Visual evaluation (observer 2)

Negative

Positive

Total (N)

Negative

Positive

Total (N)

EUROPattern (EPa) software (adjusted)

 Negative

673

2

675

620

55

675

 Positive

84

274

358

45

313

358

 Total (N)

757

276

1033

665

368

1033

Kappa agreement

0.81

 

0.79

 

Relative sensitivity* (%)

99.3

85.1

Relative specificity* (%)

88.9

93.2

  1. *To calculate relative sensitivity and specificity, the visual evaluation made by the observer (1 or 2) is regarded as correct