Table 6 Summary of method agreement (Bland–Altman plot) for the TgAb methods

From: Definition of the upper reference limit for thyroglobulin antibodies according to the National Academy of Clinical Biochemistry guidelines: comparison of eleven different automated methods

Agreement

AIA vs ARC

CEN vs ARC

CL2 vs ARC

COB vs ARC

IMM vs ARC

KRY vs ARC

LIA vs ARC

LUM vs ARC

MAG VS ARC

PHA vs ARC

No.

120

120

120

120

120

120

120

120

120

120

Bias %

(±1.96 SD)

−37

(78.7 to −152.6)

133.6

(219.2 – 48)

−115.8

(−45.1 to −186.6)

155.1

(194.3–115.9)

−54.6

(106.8–216.1)

125.8

(226.7–24.8)

141.8

(189.5–94)

125

(188.3–61.7)

156.4

(201.7–111)

147.2

(190.1–104.2)

  1. AIA AIA-2000, Tosoh Bioscience, ARC Architect, Abbott Diagnostics, CEN Advia Centaur XP, Siemens Healthineers, CI confidence intervals, CL2 AIA CL-2400, Tosoh Bioscience, COB Cobas 6000, Roche Diagnostics, e-URL experimental upper reference limit, IMM Immulite 2000 XPi, Siemens Healthineers, KRY Kryptor, Thermo Fisher Scientific BRAHMS, LIA Liaison XL, Diasorin, LUM Lumipulse G, Fujirebio, m-URL manufacturer upper reference limit, MAG Maglumi 2000 Plus, Snibe, No. number, PHA Phadia 250, Phadia AB, Thermo Fisher Scientific, r s Spearman’s rank correlation coefficient