Skip to main content

Table 1 Analytical performance characteristics of the current TgAb automated immunoassays

From: Definition of the upper reference limit for thyroglobulin antibodies according to the National Academy of Clinical Biochemistry guidelines: comparison of eleven different automated methods

Method

Immunoassay principle

Tracer/enzyme

Assay type

Imprecision (%):intra-; inter-; total

LoDd (IU/mL)

LoQd (IU/mL)

Assay range (IU/mL)

AIA

FEIA

4MUP/

Alkaline phosphatase

NC

4.3–5.1; nd;

5.5–6.0

0.12

nd

0.12–2000

ARC

CLIA

Acridinium esters

NC

1.7–6.6b; nd;

2.7–8.2b

0.07

0.31

0.07–1000

CENa

CLIA

Acridinium esters

C

2.9–5.5; 1.8–2.0;

3.5–5.8

10

30e

10–500

CL2

CLEIA

Difurat®

NC

5.1–5.5;

5.8–6.6; nd

0.005

nd

0.005–2500

COB

ECLIA

Ruthenium derivatives

C

1.3–5.6c;

2.1–8.7c; nd

10

nd

10–4000

IMM

CLIA

Adamantyl dioxetane phosphate/Alkaline phosphatase

NC

3.2–4.9;

4.6–5.8; nd

2.2

nd

20–3000

KRYa

TRACE

Europium cryptate/

XL 665

C

1.5–3.5;

6.8–20.0; nd

10

33

10–850

LIA

CLIA

Isoluminol derivatives

NC

2.3–3.2;

4.4–8.9; nd

5

10

5–5000

LUM

CLEIA

AMPPD

NC

1.8–4.6; nd;

2.5–5.3c

5.152

5.152

5.152–3000

MAG

CLIA

ABEI

NC

2.8–9.1;

5.2–9.8; nd

10

nd

10–2800

PHA

FIA

4-methyl-umbellipheryl-β-d-galactoside/β-galactosidase

NC

3.3–5.6;

2.6–6.5; nd

12

nd

12–4794

  1. 4MUP 4-methyl-umbelliferyl phosphate, ABEI N-(aminobutil)-N-(ethyl)-isoluminol, AIA AIA-2000, Tosoh Bioscience, AMPPD alkaline phosphatase-spiroadamantyl-methoxy-phosphoryloxy-phenyl-dioxetane, ARC Architect, Abbott Diagnostics, C competitive immunoassay, CEN Advia Centaur XP, Siemens Healthineers, CL2 AIA CL-2400, Tosoh Bioscience, CLIA chemiluminescence immunoassay, CLEIA chemiluminescence enzyme immunoassay, COB Cobas 6000, Roche Diagnostics, ECLIA electrochemiluminescence immunoassay, FEIA fluorescence enzyme immunoassay, FIA fluoroimmunoassay, IMM Immulite 2000 XPi, Siemens Healthineers, KRY Kryptor, Thermo Fisher Scientific BRAHMS, LIA Liaison XL, Diasorin, LUM Lumipulse G, Fujirebio, MAG Maglumi 2000 Plus, Snibe, NC non-competitive immunoassay, nd not declared, PHA Phadia 250, Phadia AB, Thermo Fisher Scientific, TRACE time resolved amplified cryptate emission
  2. aAll methods are standardized with the reference preparation MRC 65/93 and use International Units (IU/mL) except for Centaur and Kryptor which refer to a secondary standard and use Arbitrary Units (AU/mL); to obtain IU multiply for the conversion factor 2.8 (CEN) and 7.14 (KRY)
  3. bPrecision defined by the NCCLS Protocol EP5-A [26]
  4. cPrecision defined by the modified NCCLS Protocol EP5-A2 [27]
  5. dLoD and LoQ defined by the CLSI protocol EP17-A [28]
  6. eFunctional sensitivity defined as TgAb concentration with total CV ≤20%, determined for a period of two days using one lot of reagents and testing, by four instruments, multiple samples from normal patients