Table 3 Comparison of software-generated and visual pattern identification

From: Performance analysis of automated evaluation of antinuclear antibody indirect immunofluorescent tests in a routine setting

 

Visual evaluation (observer 1)

H

S

N

C

D

NM

Cyt

n = 157

n = 63

n = 37

n = 16

n = 5

n = 2

n = 59

EUROPattern (adjusted) (EPa) software

 TP

147

55

34

15

4

2

51

 FP

80

90

20

1

4

3

11

 TN

796

880

976

1016

1024

1028

963

 FN

10

8

3

1

1

0

8

 Relative sensitivity (%)

93.6

87.3

91.9

93.8

80.0

100.0

86.4

 Relative specificity (%)

90.9

90.7

98.0

99.9

99.6

99.7

98.9

 Kappa agreement

0.71

0.49

0.74

0.94

0.61

0.57

0.83

  1. H homogeneous, S speckled, N nucleolar, C centromere, D nuclear dots, NM nuclear membrane, Cyt cytoplasmic, TP true positive, FP false positive, TN true negative, FN false negative, observer 1 is considered the reference method