Table 1 The main features of the five automated immunoassay methods, as declared by the manufacturers

From: Establishment of the upper reference limit for thyroid peroxidase autoantibodies according to the guidelines proposed by the National Academy of Clinical Biochemistry: comparison of five different automated methods

System

Method

Tracer

Antigen

Imprecision (%): intra-assay; between-assay

Imprecision (%): total

LoD (IU/mL)

LoQ (IU/mL)

IMM

CLIA

Luminescence: adamantyl dioxetane phosphate

Native TPO

4.9–6.3; nr

6.1–8.1

5.0

nr

MAG

CLIA

Luminescence: N-(aminobutyl)-N-(ethyl)-isoluminol

Native TPO

6.8–7.3; 7.4–7.6

nr

0.38

nr

KRY

FIA

Fluorescence: europium cryptate/XL 665

Native TPO

2.1–7.6; 5.8–16.0

nr

1.8

9.0

PHA

FEIA

Fluorescence: β-D-galactosidase; 4-methyl-umbelliferyl-β-D-galactoside

Recombinant TPO

4.1–4.7; 2.5–4.8

nr

nr

nr

LIA

CLIA

Luminescence: isoluminol derivative

Recombinant TPO

3.6–6.2; 4.7–6.9

nr

0.6

1.0

  1. CLIA chemiluminescence immunoassay, FEIA fluorescence enzyme immunoassay, FIA fluoroimmunoassay, IMM Immulite 2000 XPi, KRY Kryptor Compact Plus, LIA Liaison XL, LoD limit of detection, LoQ limit of quantitation, MAG Maglumi 2000 Plus, nr not reported, PHA Phadia 250, TPO thyroid peroxidase